Accuris MR9620 SmartReader Multimode Microplate Reader w/ PC Software Included
Detection Modes
Absorbance, Fluorescence, Luminescence
Microplate Formats
6, 12, 24, 48, 96, and 384-well
Plate Incubation
Ambient +4°C to 45°C
- Intuitive, 10-inch color touch screen
- Stand-alone system with optional PC software (FDA CFR Part 11)
- Monochromator-based system for UV through visible absorbance
- Filter module system for fluorescence assays
- Advanced PMT detector for luminescence assays
- Optional automatic injection modules for flash luminescence
- 2 Year Manufacturer's Warranty
The Accuris SmartReader Multimode is an advanced microplate reader designed to meet the growing needs of advanced life science labs. Equipped with a full suite of high-performance optical systems, the SmartReader Multimode is your all-in-one solution for absorbance, fluorescence, and luminescence applications. The SmartReader Multimode quickly processes assays in standard 96 well plates and can also accept 6, 12, 24, 48 and 384 well plates. Programming and operation are managed from the instrument’s large color touch screen and no external computer or software is required. Optional PC software is available for easy control and data handling via a Windows™ PC. The friendly and intuitive interface guides the user through various options for assay set up and processing. The instrument features a built-in plate shaker with linear, orbital, and double-orbital shaking modes, and an incubator that can perform top heating from Ambient +4°C to 45°C. Top heating creates a temperature gradient to prevent condensation when lids are used on the microplates. Upon startup, the SmartReader Multimode performs a self-diagnostic test, followed by an automatic calibration of the entire system, ensuring consistent results across a wide range of assays.
Utilizing an advanced monochromator and a long-life, broad-spectrum xenon flash lamp, the SmartReader Multimode can produce incident light at user-selected wavelengths from 200 to 1000 nm in 1 nm steps. Processing assays is fast; a full 96 well plate can be read in under 15 seconds. The kinetic assay feature performs multiple readings over a set period, with curves for each well displayed in real-time. The optional SmartDrop plate allows fast and convenient UV absorbance evaluation of up to 16 nucleic acid and protein sample droplets. Absorbance applications include ELISA Assays, Colorimetric Assays (Bradford, Lowry, etc.), Enzyme Kinetics, Spectral Scanning, DNA & RNA Analysis, and more.
Three filter modules are included to provide the most common excitation and emission wavelengths needed for fluorescence assays. Additional filter modules are available. These advanced modules incorporate dichroic mirrors with narrow band-pass optical filters and ensure high light transmission, increased detection sensitivity, and are easily exchanged. Modules are barcoded for automatic loading of wavelength data. The SmartReader Multimode allows for automatic dynamic range selection, which allows for the gain of the photomultiplier tube to accurately capture and normalize the signal intensity of samples.
The SmartReader Multimode is capable of luminescent detection and can process both flash and glow assays. An advanced PMT (photo multiplier tube) enhances sensitivity of weak signals, prevents over-saturation of strong signals, and provides an ideal detection range for all common luminescence applications. Up to 2 sample injectors (available separately) can be installed and are compatible with both 96 and 384-well plates. Data from the SmartReader Multimode to can be exported as either a .XLS or .CSV file via the included USB flash drive, the FTP server, or by using the optional SmartReader PC Software. The SmartReader PC Software allows for remote programming & operation of the instrument via cable connection. Data from the instrument is automatically uploaded to the connected PC for further analyses. Both a Research version for academic use, and a FDA 21 CFR Part 11 compliant version are available.
The
Accuris MR9620 SmartReader Multimode Microplate Reader includes:
- Three Fluorescence Filter Cartridges (Excitation/Emission) 470/525nm, 523/564nm, & 624/692nm
- PC Software
- Power Cord
- Operator's Manual
- Manufacturer's Documentation
- 2 Year Manufacturer's Warranty
Accuris MR9620 SmartReader Multimode Microplate Reader Specifications
General Specifications
|
Measurement Technologies
|
Absorbance, Fluorescence, Luminescence
|
Plate Formats
|
6- to 384-well plates
|
Shaking
|
Linear (4.7 Hz, 7.0 Hz, 13.5 Hz), orbital and double orbital (240 – 1200 rpm)
|
Incubation Range
|
RT +4 °C to 45 °C
|
Display
|
10.0 inch touchscreen
|
Communication Ports
|
2 USB A type ports / 1 USB B type port
1 Ethernet port
RS232 interface
|
Dimensions (WxDxH)
|
16.5 x 21.7 x 15.2 in.
(42.0 x 55.0 x 38.6 cm)
|
Power Input
|
AC 100 to 240 V, 50 to 60 Hz
|
Weight
|
33 kg / 72.8 lbs
|
Absorbance Specifications
|
Light Source
|
High Energy Xenon Flash Lamp
|
Detector
|
Photodiode
|
Wavelength Accuracy
|
2 nm
|
Wavelength Repeatability
|
0.2 nm
|
Bandwidth (FWHM)
|
> 2.5 nm
|
Wavelength Range
|
200 – 1000 nm, 1nm steps
|
Absorbance
|
0.0 – 4.0 OD
|
Resolution
|
0.0001 OD
|
Accuracy
|
96-precision mode:± (1.0 %+0.003Abs)@ (0.0-2.0Abs] ;± 2.0 % @ (2.0-3.0Abs]
|
Repeatability
|
CV < 1.0 % or SD< 0.003 Fast(0.0-3.0Abs]CV < 0.5 % or SD< 0.003 Accurate(0.0-3.0Abs]
|
Stray Light
|
0.1 % @ 220 nm
|
Linearity
|
R2 > 0.999 @ [0.0-3.0Abs]
|
Measurement Speed
|
< 15 seconds
|
Fluorescence Specifications
|
Reading Mode
|
Top Reading
|
Excitation Light Source
|
High Energy Xenon Flash Lamp
|
Detector
|
Photo Multiplier Tube (PMT)
|
Detection Limit
|
≤ 1 pM
|
Wavelength Range
|
EX: 200 – 1000 nm; EM: 270 – 850 nm
|
Included Filter Cartridges
|
(EX / EM): 470 / 525nm ; 523 / 564nm ; 624 / 692nm
|
Linear Dynamic Range
|
6 logs
|
Luminescence Specifications
|
Detector
|
Photo Multiplier Tube (PMT)
|
Detection Limit
|
15 amol / well
|
Linear Dynamic Range
|
6 logs
|
Spec / Product
|
Accuris MR9600 SmartReader 96
|
Accuris MR9600-T SmartReader 96
|
Accuris MR9610 SmartReader
|
Accuris MR9611 SmartReader
|
Accuris MR9620 SmartReader
|
Measurement Technologies
|
Absorbance
|
Absorbance
|
Absorbance
|
Absorbance
|
Absorbance, Fluorescence, Luminescence
|
Wavelength Range
|
340 to 750 nm (405, 450, 492, and 630 nm filters included)
|
340 to 750nm (405, 450, 492) and 630 nm filters included)
|
200 to 1000 nm with 1 nm steps
|
200 to 1000nm
|
200 to 1000 nm, 1nm steps
|
Plate Format
|
96-Well Plates
|
96-Well Plates
|
96-well and 384-well flat bottom
|
96-well and 384-well flat bottom (cuvettes accepted in MR9611 model)
|
6- to 384-well plates
|
Test Speed
|
< 6 sec. for 96 well plate
|
<6 sec. for 96 well plate
|
96 well plate: <8 seconds (fast mode); 384 well plate: <74 seconds
|
96 well plate: <6 seconds (fast mode); 384 well plate: <74 seconds
|
< 15 seconds
|
Shaking
|
3 modes: slow, medium, fast
|
3 modes: slow, medium, fast
|
Linear; three options: (16.8Hz, 1mm), (6.68Hz, 3mm), (2.44Hz, 15mm)
|
Linear; three levels of shaking speed
|
Linear (4.7 Hz, 7.0 Hz, 13.5 Hz), orbital and double orbital (240 – 1200 rpm)
|
Incubation
|
From ambient +5°C to 50°C
|
From ambient + 5℃ to 45℃
|
From ambient + 5℃ to 45℃
|
From Ambient +5°C to 45°C
|
RT +4 °C to 45 °C
|
Light Source
|
Quartz-halogen lamp 6V/10W
|
Quartz-halogen lamp 6V/10W
|
Xenon Flash Lamp
|
Xenon Flash Lamp
|
High Energy Xenon Flash Lamp
|
|
|
|
|
|
|
|