Accuris MR9611 SmartReader Microplate Reader UV-Vis with Cuvette Port & PC Software Included
Formats Accepted
96-Well, 384-Well Flat Bottom Plates, Cuvettes
Display
10-inch High Resolution Color Touch Screen
Wavelength Range
200 - 1000 nm, 1 nm Steps
- Intuitive, 10-inch color touch screen control panel
- Accepts both 96 & 384-well plates as well as cuvettes
- Monochromator-based system
- No filters required
- Wavelength range from 200 to 1000 nm
- Stand-alone design
- No PC or special software required
- USB port for flash drive data transfer
- SmartDrop™ Accessory Plate available for DNA/RNA Quantification/Purity Analysis
The Accuris SmartReader UV-Vis Microplate Reader with Cuvette Port (Model MR9611) has been designed with ease-of-use and fast results in mind. This instrument brings extensive capabilities to the lab bench and is designed for the busy lab with multiple users. The large touch screen command center is extremely user-friendly and allows for fast set up of new protocols, as well as modifying and running existing protocols.
The monochromator-based optical system allows for precise selection of wavelengths from 200nm to 1000nm, with no filters required. This range makes the SmartReader UV-Vis useful not only for typical ELISA , Bradford Lowry, and turbidity assays but also for more advanced applications such as quantitation of nucleic acids and proteins. The SmartReader UV-Vis can be set for Endpoint, Kinetic and Spectral Scanning assays.
The MR9611 is designed for efficient processing of multiple samples and accepts standard 96 well plates, 384 well plates, as well as cuvettes for high density screening. The plate layout interface allows for a quick and intuitive setup of a plate well’s contents. Users can highlight and label multiple wells to save time. The display shows a graphical representation of the plate. Users can label each well as required for: “known standard”, “unknown sample”, “blank well” , “QC Control well”, “positive control”, “negative control”, and “empty”.
Controllable plate shaking and incubation allow for optimization of assay results. The
shaking speed can be controlled at different intensities with options for timed and pulsed
shaking before a reading, or in between readings when performing a kinetic assay. The
linear motion efficiently and uniformly mixes all samples in both 96 and 384 well plates.
Temperature can be controlled from ambient +5°C up to 45°C, allowing for the processing
of enzyme kinetics, and bacterial growth assays.
The Accuris MR9611 SmartReader Microplate Reader UV-Vis with Cuvette Port & PC Software Included includes:
- SmartReader PC Software for MR9611
- Cuvette Port
- Power Cord
- Operator's Guide
- Manufacturer's Documentation
- 2 Year Manufacturer's Warranty
Accuris MR9611 SmartReader Microplate Reader UV-Vis with Cuvette Port & PC Software Included Specifications
Model
|
Accuris SmartReader UV-Vis
|
Measurement Technologies
|
Absorbance
|
Plate Formats
|
96-well and 384-well flat bottom (cuvettes accepted in MR9611 model)
|
Display
|
10-inch color touch screen
|
Controls
|
Touch screen and mouse
|
Light Source
|
Xenon Flash Lamp
|
Wavelength Range
|
200-1000nm
|
Wavelength Precision
|
± 2 nm
|
Wavelength Accuracy
|
± 2 nm
|
Wavelength Selection
|
Monochromator
|
Read-out range
|
0 – 4.0 OD
|
Detector
|
Dual silicon photoelectric detectors for measurement and reference
|
Linearity @ 450nm
|
R2 ≥ 0.999, [0.0 – 3.0]
|
Accuracy @ 450nm
|
1.0% + 0.003 Abs (0 – 2.0]
2.0% (2.0 – 2.5 Abs]
|
Precision @ 450nm
|
CV < 0.5% or SD<0.003 (Precision mode); CV < 1.0% (Fast mode)
|
Stability @ 450nm
|
< 0.005A , (0.0 – 2.0]
< 2% (2.0 – 2.5 Abs)
|
Measurement Speed
|
96 well plate: <6 seconds (fast mode); <25 seconds (precision mode)
384 well plate: <74 seconds (only precision mode available for 384)
|
Plate Shaking
|
Linear; three levels of shaking speed
|
Incubator
|
From Ambient +5°C to 45°C
|
Temperature Uniformity
|
± 0.5°C at 37°C
|
Storage
|
16G memory, more than 10,000 test records can be stored
|
Connection Ports
|
3 USB ports, for PC, printer and USB-flash drive
|
Power Supply
|
DC24V 6.5A
|
Dimensions (W × D × H)
|
300 × 500 × 260mm / 11.8 x 19.7 x 10.25 inches
|
Weight
|
15.5 kg / 34 lb.
|
Spec / Product
|
Accuris MR9600 SmartReader 96
|
Accuris MR9600-T SmartReader 96
|
Accuris MR9610 SmartReader
|
Accuris MR9611 SmartReader
|
Accuris MR9620 SmartReader
|
Measurement Technologies
|
Absorbance
|
Absorbance
|
Absorbance
|
Absorbance
|
Absorbance, Fluorescence, Luminescence
|
Wavelength Range
|
340 to 750 nm (405, 450, 492, and 630 nm filters included)
|
340 to 750nm (405, 450, 492) and 630 nm filters included)
|
200 to 1000 nm with 1 nm steps
|
200 to 1000nm
|
200 to 1000 nm, 1nm steps
|
Plate Format
|
96-Well Plates
|
96-Well Plates
|
96-well and 384-well flat bottom
|
96-well and 384-well flat bottom (cuvettes accepted in MR9611 model)
|
6- to 384-well plates
|
Test Speed
|
< 6 sec. for 96 well plate
|
<6 sec. for 96 well plate
|
96 well plate: <8 seconds (fast mode); 384 well plate: <74 seconds
|
96 well plate: <6 seconds (fast mode); 384 well plate: <74 seconds
|
< 15 seconds
|
Shaking
|
3 modes: slow, medium, fast
|
3 modes: slow, medium, fast
|
Linear; three options: (16.8Hz, 1mm), (6.68Hz, 3mm), (2.44Hz, 15mm)
|
Linear; three levels of shaking speed
|
Linear (4.7 Hz, 7.0 Hz, 13.5 Hz), orbital and double orbital (240 – 1200 rpm)
|
Incubation
|
From ambient +5°C to 50°C
|
From ambient + 5℃ to 45℃
|
From ambient + 5℃ to 45℃
|
From Ambient +5°C to 45°C
|
RT +4 °C to 45 °C
|
Light Source
|
Quartz-halogen lamp 6V/10W
|
Quartz-halogen lamp 6V/10W
|
Xenon Flash Lamp
|
Xenon Flash Lamp
|
High Energy Xenon Flash Lamp
|
Frequently Purchased With
|
|
|
|
|
|
|