Measuring Range Max.
|
40000 J
|
Measuring Mode Dynamic 22°C
|
Yes
|
Measuring Mode Isoperibol 22°C
|
Yes
|
Measuring Mode Dynamic 30°C
|
Yes
|
Measuring Mode Isoperibol 30°C
|
Yes
|
Measurements per Hour Dynamic
|
6
|
Measurements per Hour Isoperibol
|
4
|
Reproducibility Dynamic (1g Benzoic Acid NBS39i)
|
0.15% RSD
|
Reproducibility Isoperibol (1g Benzoic Acid NBS39i)
|
0.05% RSD
|
Touchscreen
|
Yes
|
Working Temperature
|
22 to 30 °C
|
Temperature Measurement Resolution
|
0.0001 K
|
Cooling Medium Temperature
|
12 to 27 °C
|
Cooling Medium Permissible Operating Pressure
|
1.5 Bar
|
Cooling Medium
|
Tap Water
|
Type of Cooling
|
Flow
|
Flow Rate
|
60 - 70 L/H
|
Rec. Flow Rate at 18°C
|
60 L/H
|
Oxygen Operating Pressure Max.
|
40 Bar
|
Interface Scale
|
RS232
|
Interface Printer
|
USB
|
Interface PC
|
RS232
|
Interface Ext. Keyboard
|
Yes
|
Oxygen Filling
|
Yes
|
Decomposition Vessel C 6010
|
Yes
|
Decomposition Vessel C 6012
|
Yes
|
Analysis According to ASTM D240
|
Yes
|
Analysis According to ASTM D4809
|
Yes
|
Analysis According to ASTM D5865
|
Yes
|
Analysis According to ISO 1928
|
Yes
|
Analysis According to GB T213
|
Yes
|
Works According to DIN 51900
|
Yes
|
Works According to DIN EN ISO 1716
|
Yes
|
Works According to DIN EN ISO 9831
|
Yes
|
Works According to DIN EN ISO 18125
|
Yes
|
Works According to DIN EN 15170
|
Yes
|
Works According to DIN EN 15400
|
Yes
|
Works According to DIN CEN TS 14918
|
Yes
|
Works According to DIN CEN/TS 16023
|
Yes
|
Works According to DIN SPEC 19524
|
Yes
|
Permissible Ambient Temperature
|
20 to 30 °C
|
Permissible Relative Humidity
|
80 %
|
Protection Class According to DIN EN 60529
|
IP 20
|
RS 232 Interface
|
Yes
|
USB Interface
|
USB-B
|
Dimensions (W x H x D)
|
19.7 x 16.7 x 17.7 in. (500 x 425 x 450 mm)
|
Weight
|
63.9 lb. (29 kg)
|
Electrical Requirements
|
100 to 120V 50/60 Hz, 1700 W
|