The nEXT 85H pumps consist of the turbomolecular pump with a permanently attached controller containing drive electronics.
The controller controls the electrical supply to the pump with the exception of standby speed control. It has no manual controls and can only be operated through the logic interface. To operate the nEXT-85 H pump you must connect it to your own control equipment and power supply or alternatively use the Edwards TIC Turbo Instrument Controller or TIC Turbo Controller.
nEXT 85H pumps are supllied with an inlet-screen fitted into the centering O-ring for ISO version pumps and into envelope for CF version pumps. nEXT85 pumps with an NW 25 interstage-port are supplied with an inlet-strainer that fits into the interstage-port. The inlet-screen and inlet-strainer protect the pump against damage that would be caused by debris entering the pump.
The nEXT 85 H pumps have a vent-port for venting the pump and vacuum system to atmospheric pressure. The pump is supplied with a manual vent-vavle fitted; this can be replaced with a TAV5 or TAV6 solenoid-operated vent-valve. The TAV valve can be directly controlled by the on-board controller.
The nEXT pumps have a purge-port: an inert purge gas can be introduced protect the bearing and motor from corrosion. An optional vent-port adapter and purge restrictor can be fitted to the purge-port to control the flow rate of the purge gas and to filter the gas supply.
Click Here to View Product Video
The Edwards nEXT85H NW40 NW16 80W Turbomolecular Vacuum Pump includes:
- Operator's Manual
- 1 Year Manufacturer's Warranty
- Manufacturer's Documentation
Edwards nEXT85H NW40 NW16 80W Turbomolecular Vacuum Pump Specifications
Power Total Rated (W)
|
80.00 W
|
Pumping Speed Nominal
|
2820.00 l/min
|
Ultimate Pressure
|
<5x10
-9 mbar
|
Inlet Type
|
NW40
|
Pressure Backing Max.
|
13.5 Torr
|
Rotational Speed Nominal
|
1500.00 Hz
|
Ramp Up Time
|
115 s
|
Compression Ratio N2
|
>1 x 10¹¹
|
Compression Ratio He
|
4.6 x 10⁷
|
Compression Ratio H2
|
6.5 x 10⁵
|
Parameter
|
Edwards nEXT85H - CF63 NW16
|
Edwards nEXT85H ISO63 NW16
|
Edwards nEXT85H NW40 NW16
|
Power Total Rated (W)
|
80
|
80
|
80
|
Pumping Speed Nominal
|
503.99 l/min
|
503.99 l/min
|
2820.00 l/min
|
Pressure Ultimate
|
3.750 × 10⁻⁹ Torr
|
3.750 × 10⁻⁹ Torr
|
3.750 × 10⁻⁹ Torr
|
Pressure Backing Max.
|
15.75 Torr
|
15.75 Torr
|
13.5 Torr
|
Rotational Speed Nominal
|
1500.00 Hz
|
1500.00 Hz
|
1500.00 Hz
|
Ramp Up Time
|
130.00 s
|
130.00 s
|
115 s
|
Compression Ratio N2
|
1.000 × 10¹¹
|
1.000 × 10¹¹
|
>1 x 10¹¹
|
Compression Ratio He
|
2.000 × 10⁷
|
2.000 × 10⁷
|
4.6 x 10⁷
|
Compression Ratio H2
|
5.000 × 10⁵
|
5.000 × 10⁵
|
6.5 x 10⁵
|
Inlet Type
|
CF63 NW16
|
ISO63 NW16
|
NW40 NW16
|